ISO 16063-11:1999

Methods for the calibration of vibration and shock transducers — Part 11: Primary vibration calibration by laser interferometry

Methods for the calibration of vibration and shock transducers — Part 11: Primary vibration calibration by laser interferometry

Digital Document1999 edition1999Purchase Options

Referenced standards

ISO 16063-1ISO 16063-21ISO 266
Unit Price
USD120.90Save 22%
USD 155.00

Methods for the calibration of vibration and shock transducers — Part 11: Primary vibration calibration by laser interferometry

This part of ISO 16063 specifies the instrumentation and procedure to be used for primary vibration calibration of

rectilinear accelerometers (with or without amplifier) to obtain magnitude and phase lag of the complex sensitivity

by steady-state sinusoidal vibration and laser interferometry.

It is applicable to a frequency range from 1 Hz to 10 kHz and a dynamic range (amplitude) from 0,1 m/s2 to

1 000 m/s2 (frequency-dependent).

These ranges are covered with the uncertainty of measurement specified in clause 2. Calibration frequencies lower

than 1 Hz (e.g. 0,4 Hz, which is a reference frequency used in other International Standards) and acceleration

amplitudes smaller than 0,1 m/s2 (e.g. 0,004 m/s2 at 1 Hz) can be achieved using Method 3 specified in this part of

ISO 16063, in conjunction with an appropriate low-frequency vibration generator.

Method 1 (fringe-counting method) is applicable to sensitivity magnitude calibration in the frequency range 1 Hz to

800 Hz and, under special conditions, at higher frequencies (cf. clause 7). Method 2 (minimum-point method) can

be used for sensitivity magnitude calibration in the frequency range 800 Hz to 10 kHz (cf. clause 8). Method 3

(sine-approximation method) can be used for magnitude of sensitivity and phase calibration in the frequency range

1 Hz to 10 kHz (cf. clause 9).

Methods 1 and 3 provide for calibrations at fixed acceleration amplitudes at various frequencies. Method 2 requires

calibrations at fixed displacement amplitudes (acceleration amplitude varies with frequency).

Edition1999
Standard Part No11
Published Date1999
Published ByInternational Standardization Organisation
Page Count27